Fully Auto Clinical Chemistry Analyser
DC 400 |
Fully Automatic Clinical Chemistry Analyzer |
Highly-Efficient Sample Processing, Constant Speed 400 T/H |
Constant test speed of 400 T/H, optional ISE |
90 Sample Positions, 80 Reagent Positions |
90 Quartz Cuvettes |
Dry Heating Thermostatic System - Rapid Heating, Maintenance Free |
Directly Heating System Enables Faster Heating and Decreases Pre- Heating Time |
Thernostatic System by Dry Heating Keeps Reaction Solution in Cuvettes at (37.0+0.2)'C. |
Accurate Sampling System |
High-Precision Injector |
Internally and Externally Polished Sampling Probles |
Liquid Volume Tracking and Leval Detection |
Minimum Reaction Volume: 100ul |
Minimum Sampling Volumes: 1.5ul |
CV<2% |
Intelligent Online Loading System |
Reagents can be Loaded On-board During Test |
Peltier System Provides a Longer Maintenance Period to Reduce Repair Probablity. |
8-Step Warm Liquid Washing - More thorough cleaning |
Whole Probe Stero Cleaning Technology with Cross-Contamination Rate<0.05% |
Pure Water Washing in Step 1, Cleaning Solution Washing in Step 2-3, Pure Water Washing in Step 4-6, Than Wiping Module Washing in Step 7-8 |
Anti-overflow Function Prevents Liquid Spillage |
Advanced Operation System |
lntuitive user-friendly interface, easy to learn and use |
Auto sample dilution |
Auto stopping and warning when reagents or samples run out |
Calibration with linear or non-linear curves. Multi-point calibration |
Automatically detecting cuvettes status and screening flawed cuvettes |
Test Principle |
Latex Immunoturbidimetry, Immunoturbidimetry, Colorimetry |
Analysis Method |
Rate, End-point, Fixed-time |
Test Items |
Up to 78 items at the same time |
Analysis Rate |
Constant Speed 400 T/H |
Sample Tray |
90 Sample Positions (including sample, calibrator and controls) |
Sample Volume |
1.50-50.0 µL (0.1 µL increment) |
Reagent Volume |
R1: 20.0-300.0 µL |
R2/R3/R4: 0 or 20.0-300.0 µL (1 µL increment) |
|
Reagent |
Up to 4 reagents per test item |
Reagent Tray |
Reagent Positions: outer: 40 sample positions |
Reaction Volume |
100.0-360.0 µL |
Reaction Temperature |
Constant temperature control (37.0±0.2)°C |
Cuvette |
Quartz glass: 5*5mm; 90 cuvettes |
Weight |
300 kg |
Dimension |
1160 mm (L) x 790 mm (W) x 1140 mm (H) |
Water Consumption |
No more than 20 L purified water per hour |